ABSTRACT

The dissolution of nickel from nitinol alloy under different extraction conditions was investigated when biological tests of nitinol medical devices were carried out. It discussed the equivalence of these extraction conditions and found the best extraction conditions. In the experiment, two brands of nitinol were chosen and extracted under different extraction conditions according to GB/T 16886-12. Plasma emission spectrometer was used to analyse the concentration of nickel in extract liquid. The results show that there is no equivalence among these extraction conditions for nitinol materials. And it is suggested that (37 +/- 1) degree C, (72 +/- 2) h could be used as normal extraction condition and (70 +/- 2) degrees C, (24 +/- 2) h as accelerated extraction condition.


Subject(s)
Alloys , Chemistry , Biocompatible Materials , Materials Testing , Nickel , Chemistry , Temperature , Titanium